Size effect on dose output in phantoms of x-ray tubes in medical x-ray imaging.

Author: Xiaoming Zheng; Laura Nardi; Mark Murray
Publisher: Bristol : IOP Publishing, 2017.
Edition/Format:   Downloadable article : Document   Computer File : English
Publication:Biomedical physics & engineering express
Summary:
Abstract: The purpose of this work was to study how patient size affects the exponent of the power law relating dose to x-ray tube potential in clinical x-ray imaging using phantoms. Computed tomography (CT) dose phantoms of 16 cm and 32 cm in diameter were used to model children and adults respectively. A Fujifilm digital radiographic imaging system and a GE Discovery HD 750 CT scanner were employed to image both  Read more...
Rating:

(not yet rated) 0 with reviews - Be the first.

Subjects
More like this

Find a copy online

Links to this item

Find a copy in the library

&AllPage.SpinnerRetrieving; Finding libraries that hold this item...

Details

Material Type: Document, Internet resource
Document Type: Internet Resource, Article, Computer File
All Authors / Contributors: Xiaoming Zheng; Laura Nardi; Mark Murray
ISSN:2057-1976
OCLC Number: 1053705335
Notes: In: Biomedical physics & engineering express, Vol. 3, no. 6 ( 2017), p.-.
Description: 1 online resource

Abstract:

Abstract: The purpose of this work was to study how patient size affects the exponent of the power law relating dose to x-ray tube potential in clinical x-ray imaging using phantoms. Computed tomography (CT) dose phantoms of 16 cm and 32 cm in diameter were used to model children and adults respectively. A Fujifilm digital radiographic imaging system and a GE Discovery HD 750 CT scanner were employed to image both phantoms. For the Fujifilm radiographic imaging system, doses were measured at various locations of entrance, centre, exit as well as side locations of the phantoms using a PTW DIADOS diagnostic dosimeter. Standard procedure was employed for CT dose measurements for the GE CT scanner. X-ray exposures were varied by employing various tube voltages kVp and tube current-time product or mAs. This study found that the exponent in dose power law in phantoms increases as phantom size increases. Doses measured at body centres have lower uncertainties than those at x-ray exit surfaces. The power law exponent in phantoms was found to be a linear function of the depth from the entrance surface consistent with theoretical prediction and data from the literature. These results led to both kVp and body size dependent governing equations for the variations of imaging parameters in CT and radiographic imaging. These equations provide better estimates in guiding selection of optimal scan parameters in clinical x-ray imaging.

Reviews

User-contributed reviews
Retrieving GoodReads reviews...
Retrieving DOGObooks reviews...

Tags

Be the first.
Confirm this request

You may have already requested this item. Please select Ok if you would like to proceed with this request anyway.

Linked Data


Primary Entity

<http://www.worldcat.org/oclc/1053705335> # Size effect on dose output in phantoms of x-ray tubes in medical x-ray imaging.
    a schema:CreativeWork, schema:Article, schema:MediaObject, pto:Web_document ;
   library:oclcnum "1053705335" ;
   library:placeOfPublication <http://id.loc.gov/vocabulary/countries/xxk> ;
   schema:about <http://experiment.worldcat.org/entity/work/data/5448969927#Topic/biomedical_engineering> ; # Biomedical engineering
   schema:about <http://experiment.worldcat.org/entity/work/data/5448969927#Topic/biophysics> ; # Biophysics
   schema:about <http://experiment.worldcat.org/entity/work/data/5448969927#Topic/medical_physics> ; # Medical physics
   schema:about <http://experiment.worldcat.org/entity/work/data/5448969927#Topic/medical_sciences> ; # Medical sciences
   schema:contributor <http://experiment.worldcat.org/entity/work/data/5448969927#Person/murray_mark> ; # Mark Murray
   schema:contributor <http://experiment.worldcat.org/entity/work/data/5448969927#Person/nardi_laura> ; # Laura Nardi
   schema:contributor <http://experiment.worldcat.org/entity/work/data/5448969927#Person/zheng_xiaoming> ; # Xiaoming Zheng
   schema:datePublished "2017" ;
   schema:description "Abstract: The purpose of this work was to study how patient size affects the exponent of the power law relating dose to x-ray tube potential in clinical x-ray imaging using phantoms. Computed tomography (CT) dose phantoms of 16 cm and 32 cm in diameter were used to model children and adults respectively. A Fujifilm digital radiographic imaging system and a GE Discovery HD 750 CT scanner were employed to image both phantoms. For the Fujifilm radiographic imaging system, doses were measured at various locations of entrance, centre, exit as well as side locations of the phantoms using a PTW DIADOS diagnostic dosimeter. Standard procedure was employed for CT dose measurements for the GE CT scanner. X-ray exposures were varied by employing various tube voltages kVp and tube current-time product or mAs. This study found that the exponent in dose power law in phantoms increases as phantom size increases. Doses measured at body centres have lower uncertainties than those at x-ray exit surfaces. The power law exponent in phantoms was found to be a linear function of the depth from the entrance surface consistent with theoretical prediction and data from the literature. These results led to both kVp and body size dependent governing equations for the variations of imaging parameters in CT and radiographic imaging. These equations provide better estimates in guiding selection of optimal scan parameters in clinical x-ray imaging."@en ;
   schema:exampleOfWork <http://worldcat.org/entity/work/id/5448969927> ;
   schema:inLanguage "en" ;
   schema:isPartOf <http://worldcat.org/issn/2057-1976> ;
   schema:name "Size effect on dose output in phantoms of x-ray tubes in medical x-ray imaging."@en ;
   schema:productID "1053705335" ;
   schema:url <https://nls.ldls.org.uk/welcome.html?ark:/81055/vdc_100060439830.0x000001> ;
   wdrs:describedby <http://www.worldcat.org/title/-/oclc/1053705335> ;
    .


Related Entities

<http://experiment.worldcat.org/entity/work/data/5448969927#Person/murray_mark> # Mark Murray
    a schema:Person ;
   schema:familyName "Murray" ;
   schema:givenName "Mark" ;
   schema:name "Mark Murray" ;
    .

<http://experiment.worldcat.org/entity/work/data/5448969927#Person/nardi_laura> # Laura Nardi
    a schema:Person ;
   schema:familyName "Nardi" ;
   schema:givenName "Laura" ;
   schema:name "Laura Nardi" ;
    .

<http://experiment.worldcat.org/entity/work/data/5448969927#Person/zheng_xiaoming> # Xiaoming Zheng
    a schema:Person ;
   schema:familyName "Zheng" ;
   schema:givenName "Xiaoming" ;
   schema:name "Xiaoming Zheng" ;
    .

<http://experiment.worldcat.org/entity/work/data/5448969927#Topic/biomedical_engineering> # Biomedical engineering
    a schema:Intangible ;
   schema:name "Biomedical engineering"@en ;
    .

<http://worldcat.org/issn/2057-1976>
    a schema:Periodical ;
   rdfs:label "Biomedical physics & engineering express" ;
   schema:issn "2057-1976" ;
    .


Content-negotiable representations

Close Window

Please sign in to WorldCat 

Don't have an account? You can easily create a free account.